Skip to main content

Error analysis of bulk density measurements for neutron moisture gauge calibration

Allen, R.G. and Dickey, G. and Wright, J.L. and Stone, J.F. and Hunsaker, D.J. (1993) Error analysis of bulk density measurements for neutron moisture gauge calibration. pp. 1121-1126. In: Allen, R.G. and Neale, C.M.U. (eds.) Proc. Workshop ASCE Irrigation and Drainage Division. Management of Irrigation and Drainage Systems: Integrated perspectives. USA-UT-Park City, 1993/07/21-23.

[img] PDF
804.pdf

Download (327kB)

Abstract

Six bulk density sampling methods were evaluated for use in neutron gauge
calibration. All six methods provided estimates of bulk density which were
generally within 5% of bulk density profiles measured using a gamma probe.
Standard errors of estimate ranged from 3 to 7 %. When used with care,
downhole, coring, and drive samplers can be used to successfully measure soil
moisture and bulk density profiles for use in neutron probe calibration.

Item Type: Conference or Workshop Item (Paper)
NWISRL Publication Number: 0804
Subjects: Research methodology
Mass Import - autoclassified (may be erroneous)
Depositing User: Dan Stieneke
Date Deposited: 20 Nov 2010 21:55
Last Modified: 01 Dec 2016 21:33
Item ID: 868
URI: https://eprints.nwisrl.ars.usda.gov/id/eprint/868