Recirculating Purging System for Hemispherical Net Radiometers
Stevens, J.L. and Wright, J.L. (1977) Recirculating Purging System for Hemispherical Net Radiometers. Journal of Applied Meteorology. 16(5):565-566.
PDF
396.pdf Download (99kB) |
Abstract
A purging system was developed to prevent internal
condensation in hemispherical net radiometers without
frequent changing of desiccant. Continuous recirculation
of dry air through the radiometer and a desiccant
tube prevented condensate formation during cool and
wet periods.